Heated line High Fixed-tube

heatedline

The electrically heated AHLH200-PA sample tube is designed to be connected to all Ankersmid sample elements. The heated line ensures gas components in the sample stream remain above their ACID-dew point, and thereby eliminates the risk of condensation. This is a safe way to transport the sample to a heated analyzer or the special Ankersmid coolers.

Description
The AHLH200x-PA heated lines are manufactured according to the clients specification and completely confectioned in the factory at a fixed length. The sample line temperature is tp be controlled by a Pt100 temperature controller. The heater used in this type is ONE serial resistance, twisted around the tube. Due to this construction we eliminate the occurrence of cold zones or spots in the heated line, where a potential blockage could occur. The AHLH200x-PA heated lined is insulated to protect from heat loss
with flexible multi-layer glass wool. We offer a variety of standard lines, which can be fit with many options upon request. Ready-to-use with multi-pole connector.